The necessary procedures (adjustment, calibration, initialization or registration) that must be performed after parts are removed and installed, or replaced during auxiliary battery removal/installation are shown below.
*1: When performing learning using the GTS.
Click hereAudio / Visual / Telematics>PARK ASSIST / MONITORING>INTELLIGENT CLEARANCE SONAR SYSTEM>CALIBRATION
1. PRECAUTION
After turning the power switch off, waiting time may be required before disconnecting the cable from the negative (-) auxiliary battery terminal. Therefore, make sure to read the disconnecting the cable from the negative (-) auxiliary battery terminal notices before proceeding with work.
Click hereGeneral>INTRODUCTION>REPAIR INSTRUCTION>PRECAUTION
When replacing the auxiliary battery, use a new auxiliary battery of the same dimensions with a capacity of 51 Ah or more at a 20-hour rate.
2.DISCONNECT CABLE FROM NEGATIVE AUXILIARY BATTERY TERMINAL
Loosen the nut, and disconnect the cable from the negative (-) auxiliary battery terminal.
When disconnecting the cable, some systems need to be initialized after the cable is reconnected.
Click hereGeneral>INTRODUCTION>REPAIR INSTRUCTION>INITIALIZATION
3.REMOVE AUXILIARY BATTERY
Disengage the 2 claws and open the battery terminal cap in the order shown in the illustration.
| *a | Top View of Claws | - | - |
|
Push Battery Terminal Cap Here |
|
Push |
|
Open | - | - |
Loosen the nut and disconnect the cable from the positive (+) auxiliary battery terminal.
Remove the bolt.
Remove the No. 2 battery clamp as shown in the illustration.
Remove the auxiliary battery.
Remove the battery insulator from the auxiliary battery.